Soft-db Opus Suite SLM 4-Ch Module Manuale Utente Pagina 22

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 70
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 21
19
6.4 Input Sensitivity and Calibration
The input sensitivity can be manually changed in the text field or it can be calibrated using the
calibration function and a sensor calibrator. Click the button to launch the Calibration
interface.
6.4.1 Calibration Interface
Step 1 Adjust the calibration parameters
The defaults values are:
Averaging time: 5 s
Frequency: 1 kHz
Calibrator Level: 94 dB
Step 2 Install the calibrator device on the microphone
Step 3 Click START
After the average time is elapsed, the sensitivity value will update.
Step 4 Click OK to accept the sensitivity value
Vedere la pagina 21
1 2 ... 17 18 19 20 21 22 23 24 25 26 27 ... 69 70

Commenti su questo manuale

Nessun commento