Frequency (Hz)
Saturation Level
(dB)
Minimum Linear
Level (dB)
Linear Dynamic
Range (dB)
Noise Floor (dB)
12500
120.5 27.0 93.5
17.5
16000
120.5 27.6 92.9
19.5
20000
120.5 28.3 92.2
19.7
Filter Linearity (High Range)
Frequency (Hz)
Saturation Level
(dB)
Minimum Linear
Level (dB)
Linear Dynamic
Range (dB)
Noise Floor (dB)
20
132.5 51.5 81.0
7.3
25
132.5 49.2 83.3
5.3
31.5
132.5 47.1 85.4
2.0
40
132.5 44.2 88.3
7.9
50
132.5 41.8 90.7
9.3
63
132.5 39.1 93.4
9.1
80
132.5 37.1 95.4
10.6
100
132.5 32.6 99.9
10.6
125
132.5 31.4 101.1
11.6
160
132.5 31.0 101.5
11.9
200
132.5 30.7 101.8
12.7
250
132.5 30.1 102.4
12.8
315
132.5 28.7 103.8
12.9
400
132.5 28.5 104.0
13.5
500
132.5 28.2 104.3
13.4
630
132.5 27.2 105.3
13.6
800
132.5 27.0 105.5
13.6
1000
132.5 26.8 105.7
14.1
1250
132.5 27.1 105.4
14.8
1600
132.5 27.0 105.5
15.4
2000
132.5 27.4 105.1
16.2
2500
132.5 27.6 104.9
17.0
3150
132.5 28.1 104.4
18.0
4000
132.5 30.1 102.4
19.1
5000
132.5 30.4 102.1
20.1
6300
132.5 31.8 100.7
21.0
8000
132.5 32.7 99.8
22.4